How Insulin Immunoassays Actually Work
An insulin immunoassay is a laboratory test that estimates the concentration of insulin in a blood sample by measuring how much of the sample binds to specially prepared antibodies. The assay produces a number (a concentration reading in picomoles per litre) and that number is compared against reference ranges to guide clinical decision-making.
The important word is "estimate." Immunoassays work through a competitive binding reaction that is sensitive to numerous variables: the antibody batch used, the sample preparation method, the temperature and timing of the reaction, and, critically, the presence of other substances in the sample that may compete with or mimic insulin in the binding process. Every manufacturer's package insert acknowledges these limitations. The Roche Cobas platform, which is widely used in NHS clinical biochemistry laboratories, is a robust diagnostic tool designed for clinical guidance. It is not designed to produce evidence suitable for criminal proceedings without additional confirmation.
The distinction between a clinical screening test and a forensic test is not pedantic. Clinical tools are calibrated for sensitivity: they are designed to catch as many true positives as possible, accepting some false positives in exchange for not missing genuine cases. Forensic tools are calibrated for specificity and chain-of-custody rigour; they are designed to be right when they give a positive result, because a positive result in a criminal case can deprive someone of their liberty.
The Forensic Standard Versus the Clinical Standard
In the United Kingdom, the forensic standard for insulin cases in criminal proceedings runs through the Royal Surrey County Hospital laboratory at Guildford. That laboratory operates under the Forensic Science Regulator's code of practice, with validated mass-spectrometry confirmatory testing alongside any immunoassay screening, and with full chain-of-custody documentation from sample collection to reported result.
The chain-of-custody standards for forensic insulin work are specific about sample handling: the type of collection tube matters, centrifugation timing matters, storage temperature matters, and the interval between collection and processing matters. Samples handled under clinical protocols (which are optimised for speed and practicality in a hospital setting) may not meet the forensic standard even if the underlying instrument would be capable of producing forensic-quality results under different conditions.
The samples from the Letby case were processed at the Royal Liverpool clinical biochemistry laboratory, under clinical protocols. The laboratory's own 2012 protocol document explicitly states that it cannot be used to diagnose exogenous insulin on the basis of the immunoassay result alone. For a detailed examination of why the insulin screening assay and why it cannot serve as forensic proof is a significant issue in this case, that analysis maps the gap between clinical and forensic standards in full.
Why Neonates Are Uniquely Prone to False Positives
Adult insulin assay results are already subject to well-documented false-positive risk. In neonates, premature infants in particular, that risk is substantially elevated, for reasons rooted in the specific biochemistry of extremely preterm babies.
Interfering Antibodies and Cross-Reactivity
Neonates receive passive immunity from their mothers through the placenta, including a range of maternal antibodies that are not present in adults. Some of these antibodies are structurally similar enough to insulin, or to the assay's detection antibodies, to interfere with the binding reaction and produce an artificially elevated reading. This is not a theoretical concern: the clinical biochemistry literature on neonatal insulin assay interference predates the Letby case by decades. Heterophilic antibodies and anti-insulin autoantibodies are both documented sources of immunoassay interference, and their prevalence in very preterm infants is higher than in the general population.
Cross-reactivity with insulin analogues and proinsulin variants is a separate but related problem. Proinsulin (a precursor molecule produced by the same pancreatic cells as insulin) is present in elevated concentrations in stressed or sick neonates. Some immunoassay platforms partially cross-react with proinsulin, producing a reading that overstates true insulin concentration. Whether the Roche Cobas platform in use at Royal Liverpool had this cross-reactivity profile for the specific samples tested was not independently established.
The Hook Effect at Extreme Concentrations
At the upper end of the Cobas assay's range, the signal-response curve becomes non-linear. At very high concentrations of the target molecule, the detector antibodies become saturated and further additions of the target actually cause the signal to plateau or fall — the paradox known as the hook effect. This means an extremely high concentration can produce a falsely low, or even a falsely high in the opposite direction, reading compared to the true value.
The insulin result reported for one of the affected infants was numerically at a level associated with massive deliberate overdose in adult attempted-suicide cases. That is well above the physiological range in which the assay is validated to operate reliably. Professor Geoff Chase of the University of Canterbury applied physiological modelling to the prosecution's own theory about how much insulin would need to have been added to a nutrition bag to produce that reading, and concluded the numbers were not internally consistent. The hook effect and physiologically implausible readings in neonates analysis sets out the quantitative argument in full.
What Confirmatory Mass Spectrometry Would Have Shown
Mass spectrometry identifies molecules by their precise molecular mass and fragmentation pattern. It can distinguish between human insulin, synthetic insulin analogues, proinsulin, and interfering antibodies with a specificity that immunoassays cannot approach. In every other British criminal case involving an insulin allegation that reached trial with a conviction, confirmatory mass spectrometry was performed at the Guildford laboratory.
In the Letby case it was not. The prosecution proceeded on the immunoassay screening result alone. There is therefore no confirmatory evidence of the presence of exogenous insulin — only a screening result that, under the manufacturer's own guidance, requires confirmation before being treated as diagnostic.
What mass spectrometry would have shown is, in one sense, unknowable now. The samples no longer exist in a condition suitable for testing. But the absence of the test is itself significant: the step that would have been taken in every comparable case was not taken, and the result that would have confirmed or refuted the prosecution's theory was therefore never produced.
The Peer-Reviewed Literature Since 2023
The academic response to the forensic use of immunoassay insulin results in the Letby case has been substantial. Since the 2023 convictions, peer-reviewed journals in clinical biochemistry and endocrinology have published on the specific assay platform, the sample-handling conditions, and the forensic limitations of the methodology used. The published literature is consistent in the direction of its conclusions: immunoassays are screening tools; false-positive rates in neonates are non-trivial; the specific sample-handling conditions documented in this case do not meet forensic standards; and confirmatory mass spectrometry was required and was not performed.
No peer-reviewed paper has defended the forensic use of this immunoassay result under these conditions. The Joint Expert Witness Insulin Report, filed with the CCRC in May 2025 and authored by multiple specialist clinical biochemists and endocrinologists, summarises the post-conviction literature and concludes the insulin evidence cannot bear the weight placed on it in a criminal prosecution.
What a Forensic-Standard Insulin Investigation Looks Like
For contrast: a forensic-standard insulin investigation begins at the moment of sample collection. Samples are taken into tubes appropriate for forensic work, with immediate centrifugation, frozen storage, and a documented chain of custody from collection to the accredited laboratory. The Guildford laboratory then runs both the screening immunoassay and confirmatory mass spectrometry in parallel. Only if both return consistent results consistent with exogenous insulin is a forensic-level opinion provided. That opinion is peer-reviewed within the laboratory and disclosed to both prosecution and defence with full methodology documentation.
The absence of this process in the Letby insulin counts is not a minor procedural gap. It is the absence of the scientific foundation on which a forensic insulin allegation in a criminal trial is supposed to rest.